Hearing Index

Electroacoustics in 22 FDA decisions

Values are document extractions, not cross-device bench measurements. Blank cells stay unavailable.

Decisionelectroacoustic standardfreq range hzmax output db spleq input noise dbhfa full on gain dblatency msthd percentbattery hours
DEN180026 sourceANSI/CTA 2051; ANSI/ASA S3.22 sourceNot availableNot availableNot availableNot availableNot availableNot available10 source
DEN230081 sourceANSI/CTA 2051:2017 source100-10000 source105.93 source28.2 sourceNot available3.15 source1 sourceNot available
K211008 sourceANSI/ASA S3.22:2014 source<200-8000 source113 source27 source30 sourceNot availableNot availableNot available
K212609 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017; IEC 60118-13:2019 source200-8000 source120 source29 source40 sourceNot availableNot availableNot available
K213424 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source100-8700 source110 sourceNot availableNot available5.6 sourceNot available10 source
K220303 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source200-7000 source120 source24 source35 source15 source5 source100 source
K220403 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 sourceNot available114 sourceNot available45 sourceNot availableNot availableNot available
K221052 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source250-5000 source117 source32 sourceNot available14 sourceNot availableNot available
K221064 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source200-8000 source109.6 source28.5 source29.4 source5 source0.2 source8 source
K221698 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source<200-7500 source106 source30 source26 source5.7 source1 source16 source
K223137 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source200-7000 source114.4 source26.4 source45 source15 sourceNot availableNot available
K223848 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source<200-7100 source114.4 source27 source40 source15 source5 sourceNot available
K223911 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source<200-8000 source114.4 source26.4 source51 source5.8 source1 source100 source
K230538 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 sourceNot availableNot available19 source29 sourceNot availableNot availableNot available
K231550 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source200-8000 source112 source25 source42 source1 sourceNot availableNot available
K233464 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 sourceNot available114 source23 source44 sourceNot availableNot availableNot available
K233747 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051 source<200->10000 source115.2 source18.6 source48.4 source6.7 source1 sourceNot available
K243150 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017 source200-6300 source112.7 source29.9 sourceNot available8.8 source1 source6 source
K251208 sourceNot availableNot availableNot availableNot availableNot availableNot availableNot availableNot available
K254044 sourceANSI S3.22; ANSI/CTA 2051 source232-7809 source108.6 source25.2 sourceNot available9.25 source2 source3 source
K254108 sourceANSI/ASA S3.22:2014; ANSI/CTA 2051:2017; IEC 60118-0:2022 source<200->8000 source116 source21 source47 source15 source5 sourceNot available
K261187 sourceNot available100-9030 source111.9 source24.9 source31.5 source2.7 source1 sourceNot available